Specifications
Table 1-3. AccuPyc 1330 Pycnometer Specifications
Characteristic
Specification
Temperature
Stable between 15 and 35
°
C (59 to 96
°
F)
Humidity
20 to 80% relative, non-condensing
Voltage
90 to 264 VAC
Power
30 VA
Frequency
50/60 Hz
Sample Volume
0.5 - 100 cm
3
Standard Sample
Holder
19 mm OD x 39.8 mm long x 0.254 mm wall
(0.750 in. OD x 1.570 in. long
±
0.005 in. x
0.010 in. wall) for standard (10 cm
3)
holder.
Optional Sample
Modules
Provide ranges of 1 to 100 cm
3
Precision
Reproducibility typically to within
±
0.01% of
the nominal full-scale cell chamber volume.*
Reproducibility guaranteed to within
±
0.02%
of the nominal full-scale volume on clean, dry,
thermally equilibrated samples.
Accuracy
Accurate to within 0.03% of reading plus
0.03% of nominal full-scale cell chamber
volume.*
Cabinet
3ll mm W x 173 mm H x 357 mm D
(12 in. W x 7 in. H x 14 in. D)
Gases
Research grade helium is recommended. If
unavailable, use helium with a dewpoint of -67
°
C (-88
°
F) or lower.
*Nominal full-scale cell chamber volume is the sample capacity.
AccuPyc 1330
Specifications
September 1996
1-7
Содержание AccuPyc 1330
Страница 1: ...AccuPyc 1330 Pycnometer Operator s Manual V3 03 Part No 133 42808 01 January 2005 ...
Страница 6: ...CHAPTER 1 GENERAL INFORMATION Organization of the Manual Conventions Equipment Description ...
Страница 61: ...CHAPTER 5 TROUBLESHOOTING AND MAINTENANCE Troubleshooting Error and Status Messages Maintenance ...
Страница 83: ...CHAPTER 6 ORDERING INFORMATION ...
Страница 86: ...APPENDIX A CHANGING THE ANALYSIS MODULE ...
Страница 92: ...APPENDIX B ANALYSIS THEORY ...
Страница 96: ...APPENDIX C CALIBRATION THEORY ...
Страница 99: ...APPENDIX D FORMAT OF TRANSMITTED DATA ...
Страница 108: ...APPENDIX E PRINTER CHARACTER SET ...
Страница 111: ...INDEX ...