Rear Panel Connections
1.
Attach an appropriate regulator to the gas supply cylinder. Leave the
gas cylinder shut-off valves closed until instructed otherwise.
Figure 2-7. Connecting Gas Supply Cylinder to the Regulator
2.
Attach the supplied brass reducer fitting to the outlet of the regulator
shut-off valve.
3.
Turn the shut-off valve nut clockwise until it is finger tight. Then use a
9/16-in. open-ended wrench (or an adjustable wrench) to tighten the nut
an additional 1/4 turn.
Do not overtighten the fittings. Doing so can collapse the brass fitting
and cause a leak.
4.
Attach the copper tubing to the brass reducer fitting.
5.
Turn the nut on the tubing clockwise until it is finger-tight. Then use a
7/16-in. open-ended wrench (or an adjustable wrench) to tighten the nut
an additional 1/4 turn.
6.
Attach the other end of the copper tubing to the fitting on the INLET
connector on the back of the pycnometer.
Two-Stage
Pressure Reducing
Regulator
Cylinder
Shut-Off
Valve
Shut-Off
Valve Nut
Regulator
Shut-Off Valve
Brass Reducer
Fitting
Copper
Tubing
Pressure
Control Knob
Gas Cylinder
Equipment Setup
AccuPyc 1330
2-4
September 1996
Содержание AccuPyc 1330
Страница 1: ...AccuPyc 1330 Pycnometer Operator s Manual V3 03 Part No 133 42808 01 January 2005 ...
Страница 6: ...CHAPTER 1 GENERAL INFORMATION Organization of the Manual Conventions Equipment Description ...
Страница 61: ...CHAPTER 5 TROUBLESHOOTING AND MAINTENANCE Troubleshooting Error and Status Messages Maintenance ...
Страница 83: ...CHAPTER 6 ORDERING INFORMATION ...
Страница 86: ...APPENDIX A CHANGING THE ANALYSIS MODULE ...
Страница 92: ...APPENDIX B ANALYSIS THEORY ...
Страница 96: ...APPENDIX C CALIBRATION THEORY ...
Страница 99: ...APPENDIX D FORMAT OF TRANSMITTED DATA ...
Страница 108: ...APPENDIX E PRINTER CHARACTER SET ...
Страница 111: ...INDEX ...