DTA_WRN: Cal std
10% of full-scale
Cause:
The system tried to generate a calibration report, but the
calibration standard used was less than 10% of the nomi-
nal cell chamber volume. Resulting data may be inaccu-
rate.
Action:
Use a calibration standard of sufficient size (calibration
standards are available from Micromeritics). The calibra-
tion standard should occupy at least 10% of the nominal
cell chamber volume and the more nearly filled the cell
is, the better the calibration.
DTA_WRN: Volume
10% of full-scale
Cause:
The amount of sample placed in the sample cup was so
small that the percentage of precision may be inferior.
Action:
Rerun the sample using enough material to occupy the
maximum amount of the nominal cell chamber volume.
HW_ERR: A/D I/O
failure
Cause:
An error in the pressure measurement electronics has
occurred.
Action:
Call a Micromeritics Service Representative.
HW_ERR; BB RAM has
been initialized
Cause:
Battery-backed RAM has failed.
Action:
Call a Micromeritics Service Representative.
HW_ERR: Valve I/O
failure
Cause:
An error in the valve control electronics has occurred.
Action:
Call a Micromeritics Service Representative.
AccuPyc 1330
Error and Status Messages
September 1996
5-9
Содержание AccuPyc 1330
Страница 1: ...AccuPyc 1330 Pycnometer Operator s Manual V3 03 Part No 133 42808 01 January 2005 ...
Страница 6: ...CHAPTER 1 GENERAL INFORMATION Organization of the Manual Conventions Equipment Description ...
Страница 61: ...CHAPTER 5 TROUBLESHOOTING AND MAINTENANCE Troubleshooting Error and Status Messages Maintenance ...
Страница 83: ...CHAPTER 6 ORDERING INFORMATION ...
Страница 86: ...APPENDIX A CHANGING THE ANALYSIS MODULE ...
Страница 92: ...APPENDIX B ANALYSIS THEORY ...
Страница 96: ...APPENDIX C CALIBRATION THEORY ...
Страница 99: ...APPENDIX D FORMAT OF TRANSMITTED DATA ...
Страница 108: ...APPENDIX E PRINTER CHARACTER SET ...
Страница 111: ...INDEX ...