Error and Status Messages
The following error messages may appear in the display area of the pycnome-
ter. Some of the messages contain the statement: “NN/ZZ runs completed.”
When this appears, NN will be replaced with the number of runs completed
when the error occurred and ZZ will be replaced with the number of runs re-
quested.
The messages are organized into two categories: Queued Messages and Mes-
sages Displayed During an Automatic Operation.
Queued Messages
Messages placed in a queue are displayed in chronological order when you
press
. When the queue contains error messages, an asterisk is shown
next to the Reload prompt.
ANLSERR: Eq failure
NN/ZZ runs completed
Cause:
The sample being analyzed failed to equilibrate in 1000
seconds.
Action:
Check the system carefully for leaks as described in
Checking the Cell and Expansion Chambers for Leaks
later in this chapter. Make sure the sample is properly
prepared before performing an analysis.
Cause:
Sample (foams, organics, etc.) absorbs helium slowly.
Action:
Set the equilibration rate to progressively higher values un-
til reasonable equilibration times (15 sec to 120 sec) are
achieved.
ANLSERR: Fill failure
NN/ZZ runs completed
Cause:
There was insufficient pressure to allow filling within five
minutes during an analysis.
Action:
Increase the regulator pressure or, if there is insufficient
helium in the tank, obtain a new tank of helium.
CHOICE
AccuPyc 1330
Error and Status Messages
September 1996
5-5
Содержание AccuPyc 1330
Страница 1: ...AccuPyc 1330 Pycnometer Operator s Manual V3 03 Part No 133 42808 01 January 2005 ...
Страница 6: ...CHAPTER 1 GENERAL INFORMATION Organization of the Manual Conventions Equipment Description ...
Страница 61: ...CHAPTER 5 TROUBLESHOOTING AND MAINTENANCE Troubleshooting Error and Status Messages Maintenance ...
Страница 83: ...CHAPTER 6 ORDERING INFORMATION ...
Страница 86: ...APPENDIX A CHANGING THE ANALYSIS MODULE ...
Страница 92: ...APPENDIX B ANALYSIS THEORY ...
Страница 96: ...APPENDIX C CALIBRATION THEORY ...
Страница 99: ...APPENDIX D FORMAT OF TRANSMITTED DATA ...
Страница 108: ...APPENDIX E PRINTER CHARACTER SET ...
Страница 111: ...INDEX ...