10. Remove the screws securing the module mounting rails to the baseplate.
Figure A-5. Module Mounting Rails
11. Grasp the module and pull it straight up. A convenient method of removing
the module is to replace the cap completely and use the knob as a handle to
pick up the module. Observe the standoffs mounted to the baseplate and the
corresponding recesses in the rails.
12. Disconnect the cable assemblies going to the valves mounted to the block.
Figure A-6. Disconnecting Cable Assemblies
Mounting Rails
Standoffs
Disconnect Valve
Cables Here
Disconnect Electronics
Module Cables Here