![MICRO-EPSILON C-Box/2A Скачать руководство пользователя страница 89](http://html.mh-extra.com/html/micro-epsilon/c-box-2a/c-box-2a_operating-instructions-manual_1785309089.webp)
Page 89
Appendix | Control Menu
C-Box/2A
Synchronization
Synchronization
No synchronization
Sync off. The measuring rate can be freely adjusted. Value
range: from 0.4 to 80 kHz.
Internal synchronization
The C-Box/2A forms the time base.
External
synchroniza-
tion
Low-level logic (LLL)
≤ 0.7 V: Trigger not active
≥ 2.2 V: Trigger active
The synchronization signal is received from an
external signal source, e.g. function generator.
High-level logic (HLL)
≤ 3.0 V: Trigger not active
≥ 8.0 V: Trigger active
Output data rate
Output every ... measured value Value
The reduction of the output rate causes only every nth measured
value to be output. The other measured values are discarded. Any
desired averaging over n values must be set separately.
Reducing applies for following
interfaces
Analog
Ethernet
USB
Содержание C-Box/2A
Страница 95: ......