Setting Controller Parameters
C-Box/2A
Software triggering
A measured value output is started as soon as a software command is triggered. The trigger time is defined
inaccurate. After the trigger event, the sensor outputs the previously set number of readings or starts a con-
tinuous readout.
Active logic level
Logic Level Sets, at which threshold the trigger switches:
Low-level logic (LLL)
≤ 0.7 V: Low level
≥ 2.2 V: High level
High-level logic (HLL
)
≤ 3.0 V: Low level
≥ 8.0 V: High level
Number of readings
1...16382: Number of readings to be output after a trigger event
16383: Start of an infinite readout after a trigger event
0: Stop the trigger and ending the infinite measurement output
i
For all measurement tasks, it must be remembered that the combination of level / edge triggering and
external synchronization is not is possible.