Appendix | Control Menu
C-Box/2A
A 3
Control Menu
A 3.1
Tab Home
A 3.1.1
Input
Sensor 1 /
Sensor 2
Sensor status,
Sensor type, Serial
number
Selection of the connected sensor It supports sensors of ILD2300, ILD1420, ILD1750, ILD1900,
IFC2421 and IFC2422 series.
Averaging
No averaging
Measurement values are not averaged.
Moving average
About the selectable filter width N of consecutive measured values the
arithmetic mean Mgl is formed and output. Each new reading is added
and the first (oldest) reading removed from the averaging.
Recursive average
Each new metric MW (n) is weighted to the (n-1) -fold of the previous
average.
Median filter
The median is formed from a preselected filter width N of measured val-
ues. For this purpose, the incoming measured values are re-sorted after
each measurement. The mean value is then output as median. If an even
value is selected for the filter width N, then the middle two metrics are
added together and divided by two.
Laser
On / Off
Turns on or off the laser light source on the sensor.