Setting Controller Parameters
C-Box/2A
6.4
Data Recording
On the
Settings
tab, switch to the
Data recording
menu.
6.4.1
Measurement Task
Specifies which value (possibly mastered or averaged before) will be output as C-Box/2A measurement
value:
Measurement task Measuring mode
Measurement value
sensor 1
Measured value of the sensor connected
to sensor 1 connection, i.e., the C-Box/2A
value includes the value from sensor 1.
i
If you operate exclusively one sensor
on the C-Box/2A, it must be connected
to sensor 1 connection.
Thickness sensor
1-2
Forms the difference between the two
distance values of the sensors 1/2 in direct
or diffuse reflection, with two-sided distance
measurement, and outputs the result as a
thickness value.
The thickness calculation requires as
starting value the thickness of a reference
object; this value is to be defined as master
value.
Step sensor 1-2
Forms the difference between the two
distance values of the sensors 1/2 in direct
or diffuse reflection, with one-sided distance
measurement, and outputs the result as
height value.
Value C-Box/2A = Value sensor 1 minus
value sensor 2
i
The selected measuring program is used as the standard measuring program on startup.
Fields with a gray
background re-
quire a selection.
Value
Fields with dark
border require the
specification of a
value.