![MICRO-EPSILON C-Box/2A Скачать руководство пользователя страница 80](http://html.mh-extra.com/html/micro-epsilon/c-box-2a/c-box-2a_operating-instructions-manual_1785309080.webp)
Page 80
Appendix | ASCII Communication with Sensor
C-Box/2A
A 2.3.6.13 Two-point Scaling
ANALOGSCALE1 STANDARD|(TWOPOINT <Minimum Measured Value> <Maximum Measured
Value>)
Setting the scaling of analog output1.
The standard scaling is for distances -MR/2 to MR/2, for thickness measurement 0 to 2 MR (MR = measuring
range), for intensity 0 to 100 %
If the minimum and maximum measured value is ‚0‘, the standard scale is used.
The minimum and maximum measured value must be indicated in millimeters (distance/thickness) respec-
tively % (intensity).
The available output range of the analog output is then divided between the minimum and maximum mea-
sured value. The minimum and maximum measured value must be between -1024.0 and 1024.0 mm, four
decimal places.
ANALOGSCALE2 STANDARD|(TWOPOINT <Minimalum Measred Value> <Maximum Measured
Value>)
Setting the scaling of analog output2.
The standard scaling is for distances -MR/2 to MR/2, for thickness measurement 0 to 2 MR (MR = measuring
range), for intensity 0 to 100 %.
If the minimum and maximum measured value is ‚0‘, the standard scale is used.
The minimum and maximum measured value must be indicated in millimeters (distance/thickness) respec-
tively % (intensity).
The available output range of the analog output is then divided between the minimum and maximum mea-
sured value. The minimum and maximum measured value must be between -1024.0 and 1024.0 mm, four
decimal places.
Default = STANDARD
Содержание C-Box/2A
Страница 95: ......