Operation
C-Box/2A
5.4
Operation Using Ethernet
Dynamic web pages are generated in the C-Box/2A which contain the current settings of the C-Box/2A and the peripherals. The op-
eration is only possible while there is an Ethernet connection to the C-Box/2A.
5.4.1
Requirements
You need a web browser (e.g. Mozilla Firefox or Internet Explorer) on a PC with a network connection. Decide about connecting the
C-Box/2A to a network or directly to a PC.
The C-Box/2A is delivered as standard with a fixed IP address. If you do not require a static IP address, you can enable DHCP (Dy-
namic Host Configuration Protocol) as automatic IP address allocation. The controller will be assigned an IP address by the DHCP
server
.
If you have set your browser so that it accesses internet through a proxy server, please add the IP address of the controller to the IP
addresses that should not be routed through the proxy server in the settings of the browser.
Parameter
Description
Address type
Static IP address (standard) or dynamic IP address (DHCP, Standard)
IP address
Static IP address of the controller (only active if no DHCP is selected).
Gateway
Gateway to the other subnets
Subnet mask
Subnet mask of the IP subnet
Fig. 9 Basic Ethernet settings
5.4.2
Access via Ethernet
Direct connection to PC, controller with static IP (Factory setting)
Network
PC with static IP
PC with DHCP
Controller with dynamic IP, PC with DHCP
Connect the C-Box/2A („Ethernet“ female connector) with a PC via an Ether-
net direct connection (LAN). Use a LAN cable with RJ-45 connectors for this.
Connect the controller with a switch
(Intranet). Use a LAN cable with RJ-45 con-
nectors.