Functional Principle, Technical Data
C-Box/2A
Model
C-Box/2A
Analog output
- 1 current output per connected sensor
4 – 20 mA
- 1 voltage output per connected sensor; programmable:
Unipolar 0 – 5 V / Unipolar 0 – 10 V
Bipolar ± 5 V / Bipolar ± 10 V
- Tolerance of current and voltage output: 0.04 %
Laser switch off
- Switch respectively voltage input:
switching input connected with
> laser = on
switching input open
> laser = off
input voltage < 3 V (HTL)
> laser = on
input voltage > 8 V (HTL)
> laser = off
Firmware
Measurement configurations can be saved (max. 8)
two languages (English, German), can be updated
LED
For successful connection controller/sensor, Ethernet
Power supply
- 13 – 30 VDC for full functionality, power consumption max. 200 mA
without sensor
- 10 – 13 VDC with reduced DA converter function, power consumption
max. 200 mA without sensor, analog output 0 - 5 V or ± 5 V only
- Reverse polarity protection
- No galvanic isolation, all GND signals are connected internally and with
the housing
Power consumption sensors
Maximum two sensors from internal power supply
Weight
Appr. 210 g
Case dimensions
Appr. 103 x 39 x 106 mm
Protection class
IP40