Appendix | ASCII Communication with Sensor
C-Box/2A
A 2.3.1.6 Synchronization
SYNC NONE|INTERNAL|EXTERNAL [LLL | HLL]
- NONE: Sensors are not synchronized, the C-Box/2A runs with its own clock and takes just available sen-
sor values.
- INTERNAL: C-Box/2A produces Sync impulse
- EXTERNAL: External Sync impulse is looped through to the sensors
In the case of external triggering it can still be switched between Low Level Logic (LLL) and High Level
Logic (HLL).
Low Level Logic (0 ... 0,7 to 2,8 ... 30)
High Level Logic (0 ... 3 to 8 ... 30)
Default = INTERNAL LLL
A 2.3.1.7 Booting the Controller
RESET [ALL]
The C-Box/2A restarts.
- ALL: Also restart the sensors.
A 2.3.2
Triggering
A 2.3.2.1 Trigger Selection
TRIGGER NONE|EDGE|PULSE|SOFTWARE
Selection of trigger mode
- NONE: No triggering
- EDGE: Level triggering via TRG-IN (Measuring value output depends on TRIGGERCOUNT)
- PULSE: Gate triggering via TRG-IN (continuous measuring value output while TRG-In is inactive.)
- SOFTWARE: Triggering via the command TRIGGERSW (measuring value output depends on TRIGGER-
COUNT)
Default = NONE