Setting Controller Parameters
C-Box/2A
6.4.2
Measuring Rate
Measuring rate
Measuring rate
(kHz)
0.5 / 1.0 / 2.0 / 4.0
When synchronization is switched off, the
measuring rate can be set freely. Value
range: from 0.4 to 80 kHz. Otherwise, the
possible measuring rates are specified by
the connected sensors / controllers
.
Data rate Web
diagram (kHz)
0.5 / 1.0
i
The Web diagram interface uses a
slower data rate for data transmis-
sion. So for higher measuring rates
not all the measured values will be
visible in the diagram or saved to file.
Sensor / Controller Measuring rate
ILD 1420
0.25 / 0.5 / 1 / 2 / 4 kHz
ILD 1750
0.3 ... 7.5 kHz (continuously adjustable)
7.5 kHz / 5 kHz / 2.5 kHz / 1.25 kHz / 625 Hz / 300 Hz (adjustable)
ILD 1900
0.25 ... 10 kHz (continuously adjustable)
10 kHz / 8 kHz / 4 kHz / 2 kHz /1.0 kHz / 500 Hz / 250 Hz (adjustable)
ILD 2300
1.5 / 2.5 / 5 / 10 / 20 / 30 / 50 kHz. Please note that a measurement frequency of
50 kHz involves a reduction of the sensor measuring range.
confocalDT IFC2422 Continuously adjustable 6.5 kHz ... 0.1 kHz, step size 1 Hz
Fig. 18 Preset measuring rates
Fields with a gray
background re-
quire a selection.
Value
Fields with dark
border require the
specification of a
value.