Appendix | Control Menu
C-Box/2A
A 3.2.3
Processing
Filter/Averaging
Averaging
No averaging
Measurement values are not averaged.
Moving aver-
age
2 / 4 / 8 / 16 / 32 / 64 /
128 / 256 / 512
About the selectable filter width N of consecutive
measured values the arithmetic mean Mgl is formed
and output
Recursive
average
2 / 4 / 8 / 16 / 32 / 64 /
128 / 256 / 512 / 1024
/ 2048 / 4096 / 8192 /
16384 / 32768
Each new metric MW (n) is weighted to the (n-1)
-fold of the previous average
Median filter
3 / 5 / 7 / 9
The median is formed from a preselected filter width
N of measured values
Mastering/Zeroing Mastering is Master value (mm)
disabled
Value
Set master value
Trigger zeroing or mastering
Master value range:
-1024 to 1024 mm.
enabled
Reset master value
Cancel zeroing or mastering
Trigger mode
Selected mode
No triggering
Level triggering
Value
Edge triggering
Software triggering