Chapter 3 - Installing Instrument Boards
75
Setting the DIP Switches
The procedure for determining the DIP switch settings is explained
below. As a general rule note no two carrier boards can have the
same combination of DIP switch settings.
1. Set the board base address:
a) Start by converting the hexadecimal address into binary.
Hexadecimal Binary
1 0001
2 0010
3 0011
4 0100
5 0101
6 0110
7 0111
8 1000
9 1001
A 1010
B 1011
C 1100
D 1101
E 1110
F 1111
Table 3-13: Hexadecimal to binary conversion table
b) Use the binary value to set the switches (0 = on, 1 = off),
starting at A0 on the right and working toward A15 on the
left, using as many switches as necessary.