Problem
Possible Causes
Solutions
2. Locate the leak and correct.
CO
2
out of range alarm 1. Capnograph reading is incorrect.
1. Calibrate or replace the capnograph.
IRMA out of range
alarm
1. Operating ambient pressure is out of range
2. The internal operating temperature is out of range.
1. Operate the capnograph under
environmental conditions stated in the technical
specification.
2. Replace capnography sensor.
IRMA zero required
alarm
1. Capnograph has lost calibration
1. Calibrate the capnograph.
Hight Temperature
alarm
1. The ambient temperature is above 50 ° C.
1. Operate the ventilator under environmental
conditions stated in the technical specification.
Low Temperature alarm 1. The ambient temperature is below -18 ° C.
1. Operate the ventilator under environmental
conditions stated in the technical specification.
SpO
2
Sensor alarm
1. The SpO2 sensor is connected to the equipment but
is off the finger.
1. Connect the oximeter to the patient's finger.
Check SpO
2
alarm
1. The oximeter has no sensor attached;
2. The connected sensor is defective;
3. Interference detected;
4. Too much of ambient light;
5. Unknown sensor.
1. Connect an oximetry sensor.
2. Replace the oximetry sensor.
3. Use the oximeter in an environment without
interference.
4. Decrease the ambient light.
5. Replace the oximetry sensor.
Check cable alarm
1. The oximeter cable is disconnected.
1. Connect the oximeter sensor cable.
Low Perfusion alarm
1. The monitored perfusion is below the set alarm limit.
1. Change the ventilation parameters or set the
lower limit of the Perfusion alarm.
Looking for pulse alarm 1. The oximeter is looking for pulse
1. Wait for the oximeter to find the patient's
pulse.
Activating SpO
2
alarm
1. The oximeter being activated.
1. Wait for the oximeter to activate.
SpO
2
demo alarm
1. The oximeter is generating a demonstrative curve.
1. Disconnect and reconnect the oximetry
sensor or replace the sensor.