84
OxyMag_rev26
11.
Preventive Maintenance
Caution
•
Oxymag should have their maintenance performed only by a qualified technician, trained and authorized by
MAGNAMED. Failure to comply will void the manufacturer's warranty and obligations regarding the ventilator.
•
Failure to perform maintenance could affect the safety and performance of the ventilator.
•
All maintenance shall be done with the patient disconnected from the equipment.
11.1
Indication of the need for periodic maintenance
The equipment displays in the home screen the preventive maintenance symbol
when 5000 hours or more had passed since
the last maintenance.
11.2
Daily checks and/or prior to use
•
Cleaning the equipment;
•
Integrity of the power cord of the AC/DC converter;
•
Correct operation of the visual and audible alarm
system;
•
Installation and cleaning of filters;
•
Correct display of the screen;
•
Correct use of touch screen;
•
Full battery;
•
Correct operation of the equipment panel keys;
•
Correct operation of the knob and confirm button;
•
Correct installation of the breathing circuit (including
the existence of diaphragm of the exhalation valve).
•
Daily check should be performed with the patient disconnected.
11.3
Internal Lithium Battery
This battery is Responsible to power the equipment in the absence of electricity and its duration in normal operation is specified in
15.3.1 Electrical Characteristics.
•
This equipment must ALWAYS remain connected to the mains so that there is sufficient
charge during a power outage.