Cleanning and Sterilization
10.1
Equipment cleaning
External ventilator surfaces
External ventilator surfaces of Oxymag should be cleaned with
a clean, soft cloth moistened with the enzymatic detergent.
Caution
•
Ensure that no residue builds up in the equipment connections.
•
For cleaning do not use non-compliant products to polymer
•
For touch screen cleaning avoid:
o
Use substances other than glass cleaner and do not use solutions vinegar base.
o
Use the dirty cloths and handle the display carefully.
•
Not to be used for cleaning or disinfecting the phenol (> 5%) ketone, formaldehyde, hypochlorite,
chlorinated hydrocarbons, aromatic hydrocarbons, inorganic acids, and quaternary ammonium
compounds.
Respiratory circuit, proximal flow sensor and exhalation valve
The components that come in directly contact with respiratory gases shall be periodically disassembled in components, as
accessory list, for cleaning, disinfection or sterilization.
Circuits and parts of silicone should be cleaned according the following steps:
10.1.2.1
Wash
a) Always use potable water for this procedure;
b) Use neutral and enzymatic detergent. Dilution should be performed as recommended by the manufacturer.
c) Immerse the entire body of the flow sensor and the silicone line in the detergent solution, keeping the solution in contact
with the accessories for at least 3 minutes;
d) The external parts of the parts should be cleaned with a clean, soft cloth moistened with the enzymatic detergent. The
internal parts must be cleaned by immersion.
10.1.2.2
Rinse
a) Always use potable water for rinsing;
b) Thoroughly rinse the external surface of the accessories with potable water.
c) Rinse the internal surface by injecting potable water under pressure at least 5 times.