Specifications for Maintenance and Calibration
Caution
•
The processing time is after stabilizing at the specified temperature and pressure.
•
Check the efficiency of sterilization through chemical or biological indicators.
Table 38: Specification for maintenance and calibration
Item
Description
Specification
Tolerance
Unit
1
Review and REPLACEMENT OF
DIAPHRAGMA (3800248)
50 cycles of steam
sterilization
---
Cycles
10,000 h or 2 years
500
Hours
2
Review and REPLACEMENT OF O
2
CELL (3902020)
10,000 h or 2 years
500
Hours
3
Review and REPLACEMENT OF
BATTERY (2702236)
10,000 h or 2 years
500
Hours
4
Exhalation Valve (3200251)
50 cycles of steam
sterilization
---
1 month
Cycles
months
5
Breathing Systems Adult (1703218),
Pediatric (1702654)
6
Universal Connector with silicon line
1,3 m