15
OxyMag_rev26
2.4
Components of transport ventilator
Figure 1: Frontal Panel of transport ventilator.
Figure 2: Side Keyboard
Table 3: Description of the frontal panel and the side keyboard components of the transport ventilator
Components of Figures 1 and 2
1. LIQUID CRYSTAL DISPLAY WITH TOUCH SCREEN
Visual and graphical presentation of the setting parameters with touch screen.
2. KNOB BUTTON
This button is used for most of the adjustments to be made in the Transport Ventilator Oxymag.
•
Select the parameters to be set on the display by directly touching the corresponding button;
•
The selected button will change color to YELLOW allowing the change of values or adjustments;
•
Set the desired value by turning the knob clockwise or counter-clockwise;
•
To confirm press the button
When the button returns to its original color the parameter set will be in effect.
3. KEYBOARD
The keypad buttons allow quick access to the ventilator functions.
4. HANDLE
This handle allows to carry the ventilator during rescue and emergency operations. At the rear of the ventilator, there is a support that can be easily
adapted to the patient stretcher.
5. ALARM INDICATOR LIGHT
–
RED
The alarm indicator light flashes when an alarm condition of high priority occurs. When in silent mode, it remains activated indicating the alarm
condition.
6. LOCK TOUCH SCREEN
This key allows to lock or unlock the touch screen. When the commands on the display are locked, press this key for 2 seconds to release
them. To lock again simply press this key once or wait 60 seconds without touching the screen.
7. GREEN LED
–
CONNECTION TO MAINS
The GREEN LED will bel it when the DC power inlet or power supply 12VDC inlet are connected