30
OxyMag_rev26
Note
•
The selection of the patient type on startup will perform the initial setting of the transport ventilator and release
certain ventilation modes.
•
There are three types of flow sensors
o
NEO
–
Neonatal
–
Range from -20 to +20 L.min
-1
o
PED
–
Pediatric
–
Range from -50 to + 50 L.min
-1
o
ADU
–
Adult
–
Range from -150 to +150 L.min
-1
•
Body Mass Index Formula:
Weight [kg]
IMC = ------------------------
(Height [m])
2
Normal Startup Sequence
1.
Initial screen of Oxymag
–
Turn on the ventilator through on-off switch on the left side of the equipment. By turning on, note
that a triple “beep” goes off in conjunction with the light alarm indicator, meaning that the audible and visual alarm are
operational.
•
If you do not hear a triple “BEEP” sound or do not see the light alarm indicator flashing, avoid
the use of the equipment, because there will be no audible or visual alarm indication.
2.
Press the key
corresponding to the type of patient to be ventilated and connect the flow sensor indicated in the patient’s
breathing circuit. The ventilator will be initialized in the mode indicated in
: Modes available for the types of patient
3.
Last Set Button
–
This button brings back the last parameters set and saved when the equipment was switched off for the last
time. This saving is automatic (optional).
4.
By pressing the key NEONATAL the ventilator will start ventilation with the following parameters: