96
OxyMag_rev26
Item
Parameter
Specification
Resolution
Unit
1
Tidal Volume
20 to 2500
100 to 2500: 10
mL
20 to 100: 5
2
Respiratory rate
0 to 150
(2)
1
min
-1
3
Rise time
0 to 2,0
0,1
s
4
Pause
0 to 70
10
%
5
Maximum Limit Pressure
0 to 60
1
cmH
2
O
6
Inspiratory Pressure
1 to 60
1
cmH
2
O
7
Delta of support pressure (∆PS)
OFF; 5 to 60
1
cmH
2
O
8
PEEP
0 to 40
1
cmH
2
O
9
Assisted Sensitivity (Pressure)
OFF; -0,2 to -10
-0,2 to -2,0:- 0,2
cmH
2
O
-2 to -10:- 1
10
Assisted Sensitivity (Flow)
OFF; 0,5 to 30,0
0,5
L.min
-1
11
Automatic Inspiratory Flow
(3)
0 to 150
1
L.min
-1
12
Inspiratory Flow (Neonatal)
4 to 20
1
L.min
-1
13
Cycling by Flow at Pressure
Support
5 to 80
5
%
14
O
2
Concentration
35 to 100
1
%
21 to 100 (with blender)
1
%
15
Inspiratory Time
0,1 to 10
0,1 to 0,7:0,01
s
0,7 to 1:0,05
1 to 10:0,1
16
Inspiratory Flow Waveform
Square,
Decelerated,
Accelerated,
Sine
---
---
17
CPAP
(4)
1 to 40
1
cmH
2
O
18
High Pressure
5 to 55
1
cmH
2
O
19
Low Pressure
0 to 40
1
cmH
2
O