100
OxyMag_rev26
Item
Parameter
Range
Resolution
Tolerance
Unit
14
Minute volume
(Sensor Pediatric
–
PED)
(9)
0,01 to 50,0
0,001
(0.10L or 3%
value
measured)
(4)
L
15
Minute volume
(Sensor Neonatal
–
NEO)
(9)
0,001 to
20,0
0,001
(0.06L or 3%
value
measured)
(4)
L
16
Tidal Volume Inspired
(3)
0,001 to
3000
0,01 to
3000:10
( 5% value
measured)
mL
18
Inspiratory Time
0,05 to 60,0
0,01
0,01s
s
19
Expiratory Time
0,05 to 60,0
0,01
0,01s
s
20
Ratio I:E
1:100,0 to
100,0:1
1:0,1
2%
---
21
Respiratory Rate
0 to 200
1
(1min
-1
or 1%
value measured)
min
-1
22
Airway Resistance
–
R
AW
0 to 200
1
1
cmH
2
O/L/
s
23
Dynamic Compliance
(C.Dyn)
0 to 200
0,1
1 mL.cmH2O
-1
or 10% value
measured
mL.cmH
2
O
-1
24
Static Compliance
(C.Stat)
0 to 200
0,1
1 mL.cmH2O
-1
or 10% value
measured
mL.cmH
2
O
-1
25
FiO
2
(Oxygen
Concentration)
12 to 110
0,1
±(1% in volume or
2% reading)
%O
2
26
Flow (flowmeter)
0 to 20
0,1
± (0.2L.min-1 or
5% reading)
L.min
-1
27
Regulated Pressure
0 to 150
1
3.75 psi
psi
28
Oxygen Consumption
(Cos. O
2
)
0 to 160
0,1
±(1 L/min or 10%
reading)
L/min
29
SpO
2
(10)
(LNCS DC-I
and LNCS YI)
70 to 100
1
± (2% non-
motion)
± (3% motion)
%
± 2% (low
perfision)
30
FC
(11)
(LNCS DC-I
and LNCS YI)
0 to 240
1
± (3 bpm non-
motion)
± (5 bpm motion)
bpm