70
OxyMag_rev26
MEDIUM PRIORITY
Delay Time
Description
High Minute volume
< 3 cycles
It is triggered when the minute volume of the patient has exceeded the alarm
value set as upper limit of minute volume
Low Minute volume
< 3 cycles
It is triggered when the minute volume of the patient is below the alarm value
set as lower limit of minute volume
High Rate
< 3 cycles
It is triggered when the respiratory rate of the patient has exceeded the alarm
value set as upper limit of the respiratory rate
Low Rate
< 3 cycles
It is triggered when the respiratory rate of the patient is below the alarm value
set as lower limit of the respiratory rate
High PEEP
< 3 cycles
It is triggered when the pressure in the end of exhalation (PEEP) has exceeded
the alarm value set as upper limit of PEEP
Low PEEP
< 3 cycles
It is triggered when the pressure in the end of exhalation (PEEP) is below the
alarm value set as lower limit of PEEP
CO
2
out of range
< 3 seconds
It is triggered when IRMA CO
2
reading is incorrect
IRMA out of range
< 3 seconds
It is triggered in one of the situations with the CO
2
sensor:
•
Internal temperature of operation is out of scale or
•
Ambient pressure of operation is out of scale
IRMA zero required
< 3 seconds
It is triggered when there is need of Zero calibration of IRMA CO
2
sensor
Hight Temperature
< 3 seconds
It is triggered when the environment condition is above 50ºC.
Low Temperature
< 3 seconds
It is triggered when the environment condition is below -18°C.
High FiO
2
< 3 cycles
It is triggered when the measured FiO
2
has exceeded the alarm value set as the
upper limit of FiO
2
Low FiO
2
< 3 cycles
It is triggered when the measured FiO
2
has exceeded the alarm value set as the
lower limit of FiO
2