69
OxyMag_rev26
HIGH PRIORITY
Delay Time
Description
High Volume
< 3 cycles
It is triggered when the measured volume exceeded the alarm value set as
the upper limit of volume
Low Volume
< 3 cycles
It is triggered when the measured volume exceeded the alarm value set as
the lower limit of volume
High EtCO
2
< 3 seconds
It is triggered when the expired CO
2
has exceeded the alarm value set as the
upper limit of EtCO
2
Low EtCO
2
< 3 seconds
It is triggered when the expired CO
2
is below the alarm value set as the lower
limit of EtCO
2
CO
2
i
< 3 seconds
It is triggered when the inspired CO
2
has exceeded the alarm value set as the
upper limit of CO
2
i
High FC
< 3 seconds
It is triggered when the heart rate has exceeded the alarm value set as upper
limit of FC
Low FC
< 3 seconds
It is triggered when the heart rate is below the alarm value set as lower limit of
FC
Low SpO
2
< 3 seconds
It is triggered when the Oxygen saturation is below the alarm value set as
lower limit of SpO
2
IRMA Adapter
< 3 seconds
It is triggered when there is one of the conditions: the IRMA adapter of CO
2
sensor is not connected or must be replaced
Restart IRMA
< 3 seconds
It is triggered when IRMA CO
2
sensor must be disconnected and re-connected
Change IRMA
< 3 seconds
It is triggered when IRMA CO
2
sensor must be replaced.
FiO2 below 18%
< 3 cycles
It is triggered when FiO2 measured is lower than 18%