Pressure support (ΔPS) is a value above PEEP and can be adjusted b 5 cmH2O and PMAX
- PEEP.
•
To obtain CPAP mode with backup ventilation, select option CPAP/PSV, set
ΔPS=OFF and select backup
ventilation.
1
and
2
Represent spontaneous cycles with pressure support at ZERO.
3
,
4
and
5
Represent spontaneous breathing cycles of the patient with pressure support different from zero. TRISE
TIME of pressure support may be adjusted so that initial flow is smoothed.
6
If the patient enters in apnea, after TAPNEA (s) the ventilator will show this condition with alarm in its message area
and alarms on screen and will initiate backup ventilation selected as settings and parameters programmed.
•
Apnea alarm should be set to a safe value for the patient. However, apnea alarm can be
RESET, in this condition, there will be no information or alarm for apnea condition and
no backup ventilation in action. The equipment operator must be aware of the Apnea
Alarm condition DEACTIVATED (OFF INDICATING ON DISPLAY).
•
If backup ventilation selected is NO BACK-UP, the equipment operator must be aware of
this situation (INDICATING ON DISPLAY).
•
Default values are only for initial reference. Reset the ventilation parameters as needed
by the patient.
Note