79
OxyMag_rev26
9.3.1.3
Minute volume alarm (MV)
To test the operation of the high minute volume alarm, set the upper limit of the minute volume alarm so that it is less than the
minute volume monitored on the unit. To test the operation of the low minute volume alarm, set the lower limit of the minute
volume alarm so that it is greater than the minute volume monitored on the unit.
9.3.1.4
Respiratory Rate alarm
To test the operation of the high respiratory rate alarm, set the upper limit of the respiratory rate alarm so that it is lower than
the monitored respiratory rate on the device. To test the operation of the low respiratory rate alarm, set the lower limit of the
respiratory rate alarm so that it is higher than the respiratory rate monitored on the device.
9.3.1.5
Volume alarm
To test the high volume alarm, set the upper limit of the volume alarm so that it is lower than the tidal volume set on the device.
To test the low volume alarm, set the lower limit of the volume alarm so that it is higher than the inspiratory volume monitored
on the device.
9.3.1.6
FiO2 alarm
Para testar o alarme de concentração de O2 alto, ajustar nos alarmes uma concentração máxima abaixo do ajustado na
modalidade. Para testar o alarme de concentração de O2 baixo, ajustar nos alarmes uma concentração mínima acima do
ajustado na modalidade.
9.3.1.7
EtCO2 alarm
Para testar o alarme de EtCO2, é necessário que um capnógrafo esteja conectado ao ventilador. Monte o circuito completo
para uso com capnógrafo e assopre o adaptador de vias aéreas para que o ventilador monitore os parâmetros de capnografia.
Para testar o alarme de EtCO2 alto, ajuste o limite superior do alarme de EtCO2 de forma que este seja menor que a EtCO2
monitorada no aparelho. Para testar o alarme de EtCO2 baixo, ajuste o limite inferior do alarme de EtCO2 de forma que este
seja maior que a EtCO2 monitorada no aparelho.
9.3.1.8
Pulse rate alarm
To test the heart rate alarm, a pulse oximeter must be connected to the ventilator. Mount the circuit and connect the oximeter
to your finger so that the ventilator monitors the oximetry parameters. To test the high heart rate alarm, set the upper limit of
the heart rate alarm so that it is lower than the heart rate monitored on the device. To test the low heart rate alarm, set the lower
limit of the heart rate alarm so that it is higher than the heart rate monitored on the device.
9.3.1.9
CO2 alarm
Para testar o alarme de CO2 inspirado, é necessário que um capnógrafo esteja conectado ao ventilador. Monte o circuito
completo para uso com capnógrafo e assopre o adaptador de vias aéreas para que o ventilador monitore os parâmetros de
capnografia. Ajuste o alarme de CO2 inspirado de forma que este seja menor que o CO2 inspirado monitorado no aparelho.
9.3.1.10
SpO2 alarm