65
OxyMag_rev26
1
Represents a spontaneous cycle without pressure support at P.Low (Low Continuous Airway Pressure);
2
Represents a breathing cycle with Pressure Support assist (above P.Low);
3
to
4
Represents a synchronized transition to P.High (High Continuous Airway Pressure);
5
Represents a transition from P. High to P. Low synchronized;
Transitions of levels from P.Low P.High or P.High P.Low occur in the final quarter of T.Low and T.High, respectively, by synchronizing
with the effort of the patient. The information of what trigger activated the inspiratory cycle is reported in the message area and screen
status.
•
Default values are only for initial reference. Reset the ventilation parameters as needed by
the patient.
Note
•
Respiratory rate monitored is result of the spontaneous breathing of the patient
•
Pressure support (ΔPS) is a value above P.High or P.Low and can be adjusted between 5
cmH2O and PMAX
- PHigh.
•
Changes in pressure levels are synchronized.