AC mains supply with frequency values 5% below or 5% above nominal may result in
switching to the internal battery, but the equipment will maintain its normal operation.
Caution
•
The temperature alarm will be triggered if the environmental condition is below -10ºC or above 50ºC (low/high
temperature alarm
–
medium priority).
Ventilation Modes
Table 27: Ventilation modes
Modes
(1)(2)(3)
Description
Mode in Apnea
(BACKUP)
(4)
VCV
Volume-Controlled Ventilation
AUTO
PCV
Pressure-Controlled Ventilation
AUTO
PLV
Pressure Limited Ventilation Cycled by Time for ventilator in
neonatal setting (may have assisted cycles)
AUTO
V-SIMV + PS
Volume Controlled Synchronized Intermittent Mandatory
Ventilation with Pressure Support
IMV
–
Volume
Controlled
Intermittent Mandatory Ventilation
P-SIMV + PS
Pressure Controlled Synchronized Intermittent Mandatory
Ventilation with Pressure Support
IMV
–
Pressure Controlled
Intermittent Mandatory Ventilation
DualPAP
(5)
Bi-level Positive Airway Continuous pressure with Pressure
Support
VCV, PCV (adult and pediatric) /
PLV (neonatal), OFF.
Programmable by the Operator
CPAP/PSV
Continuous Positive Airway Pressure Ventilation with
Pressure Support
VCV, PCV (adult and pediatric) /
PLV (neonatal), OFF.
Programmable by the Operator
(1)
Non-Invasive Ventilation by Mask can be activated in all ventilation modes and, in this case, there is compensation for leaks.
(2)
Automatic compensation of compliance and small leaks in the breathing circuit.
(3)
When the ventilator enters in Neonatal mo
de (IBW ≤ 6.0 Kg) only PLV, P
-SIMV, CPAP/ PSV, DualPAP modes will be available
(4)
For the modes where backup ventilator is defined as “Auto”, always when the set apnea time is reached, the ventilator start
one ventilator cycle, witch configuration is
based on current ventilator mode.
(5)
APRV (Airway Pressure Release Ventilation) mode can be obtained by DualPAP mode with appropriate adjustment of the times and pressures.
Setting Specifications of the Ventilation Parameters
Table 28: Setting specifications of parameters.