Components of Figures 1 and 2
8. HOLD KEY (PAUSE)
This key allows to suspend inspiration maneuvers, often used in cases of chest X-ray and maneuvers to extend the time of expiration (extend the
expiration time).
If pressed during the inspiratory time of the respiratory cycle, the inspiration will be prolonged for 5 seconds; after this period, parameter Cest will be
displayed in the monitoring area at the top center of the screen. If this key is pressed during the exhalation time, expiration will be extended for 5
seconds; after this period, the parameter PEEPi will be displayed in the monitoring area in the top center of the screen.
The parameters displayed after actuating this key will be visible for 5 seconds; after this period, the monitoring upper are will again display the
parameter previously displayed.
9. MANUAL KEY
This key triggers an inspiratory cycle of support pressure. And is active in the VCV, PCV, V-SIMV, CPAP/PSV, P-SIMV, DUALPAP, PLV modes.
10. O2 100% KEY
By pressing key to “O2 100%” the oxygen concentration will remain at 100% during the next 90 seconds. This feature can be use
d for pre-aspiration
and post-aspiration procedures of secretions from the airways.
When the ventilator is in STAND-BY, after pressing this button for 2 seconds, an oxygen flow meter will be displayed in the ventilator screen. At the
top of the screen, the set value of the flow meter is displayed and at the bottom the value measured of the flow delivered is displayed. Use the button
“Set and Confirm” to change the value of the desired flow.
11. FREEZE KEY
Freezes the graph layout to allow the analysis of the curves.
Figure 3: Right side view
Figure 4: Back view
Table 4: Description of components of the right and the back side of the transport ventilator.
Description of the items in Figure 3 and 4
12. FLOW SENSOR CONNECTIONS
These connectors are used to connect the flow sensor.
15. CONNECTOR TO THE CO2 SENSOR or SpO2 SENSOR
Connection to the CO2 sensor of PHASE IN or MASIMO SpO2 sensor
(these sensors are optional)
13.
CONNECTOR 22M/15F OF INSPIRATORY FLOW
Standard connection of the inspiratory flow to the breathing circuit of the
patient
16. PLUG OF POWER SUPPLY
–
BASE
This plug is used together with the support base of Oxymag and the power
supply system. The inlet is +12VDC. Note the polarity indication and the
recorded voltage DC power.
14 EXHALATION VALVE
Connection of the expiratory limb of the breathing circuit of the patient.
17. LABELING TAG
This is the INMETRO seal of compliance and safety.