![Laversab 6600-NG Скачать руководство пользователя страница 89](http://html1.mh-extra.com/html/laversab/6600-ng/6600-ng_user-manual_3905006089.webp)
6600-NG User’s Manual
125-9110A
Page 89
Command
Description
Data sent from
PC
Data sent from
6600-NG
PM=n
Set the Pt mode to n, per the following
n
Pt mode
1
Measure
2
Leak
3
Control
n
0 if OK; or
Error code
SM
Read the current Ps1 mode
None
n=1 to 3
SM=n
Set the Ps1 mode to n, per the following
n
Ps1 mode
1
Measure
2
Leak
3
Control
n
0 if OK; or
Error code
DM
Read the current Ps2 mode
None
n=1 to 3
DM=n
Set the Ps2 mode to n, per the following
n
Ps2 mode
1
Measure
2
Leak
3
Control
n
0 if OK; or
Error code
PJ=n
Jog the Pt Target value Up (n=0) or Down (n=1)
n
0 if OK; or
Error code
SJ=n
Jog the Ps1 Target value Up (n=0) or Down (n=1)
n
0 if OK; or
Error code
DJ=n
Jog the Ps2 Target value Up (n=0) or Down (n=1)
n
0 if OK; or
Error code
KJ=n
Jog the Pt Rate Target value Up (n=0) or Down
(n=1)
n
0 if OK; or
Error code
RJ=n
Jog the Ps1 Rate Target value Up (n=0) or Down
(n=1)
n
0 if OK; or
Error code
2J=n
Jog the Ps2 Rate Target value Up (n=0) or Down
(n=1)
n
0 if OK; or
Error code
Lm
Read the current limit value for Limit m, where m
defines the type of limit per the table below.
m
Limit type
1
Max knots
2
Min knots
3
Max knots/min
4
Max Mach
5
Max Feet
6
Min Feet
7
Max Feet/min
8
Max Dfin
None
Value for Limit