6600-NG User’s Manual
125-9110A
Page 76
Figure 6.1 : Calibration (invalid setup)
If the setup is done correctly, selecting option ‘
8
’ on the Functions menu will bring up the screen shown
below.
Figure 6.2 : Calibration (password)
Details of the information provided on the Lower section of this screen are provided in Section 3.10.8.
To display the main calibration screen, enter the password followed by
ENTER
. The main calibration
screen will appear as follows.
0.0008
0.0000
Pt inHg
Pt
Ps2
Ps1
Measure
knots/min
0.000
4.000
inHg
inHg
29.9225
29.9219
Qc inHg
30.0000
30.0000
inHg/min
inHg/min
0.000
0.000
4.000
4.000
Measure
Measure
AC
92%
Calibration
Total operating time: 1.5 hours
Cal. Dates (mm/dd/yy) -- Last cal: 09/01/16 Next cal due:09/01/17
Please enter calibration password:
Press
CANC
to exit
During calibration, all 3 outputs must be in Measure mode
and units for each output must be as follows:
Pt: Qc inHg ; Ps2: inHg ; Ps1: inHg
Press
CANC
to exit