6600-NG User’s Manual
125-9110A
Page 52
Figure 3.38 : Simultaneous Pt , Ps2 & Ps1 leak-check (after 3-min)
When
CANC
is pressed, all three outputs will return to Control mode.
3.7
SELF-TEST FUNCTION
The Self-Test must not be performed when the tester is connected to the aircraft. Therefore, it is up to the
user to run the Self-Test after ensuring that all three outputs are open to ambient (close to “Ground”) and
all three outputs are in Measure mode.
Note: The Self-Test should be performed each time the tester is to be used on an aircraft, just before
connecting to the aircraft.
To start the Self-Test, press
TEST
. If the conditions for running a Self-Test are not met, the following
message will be displayed in the Lower section.
Figure 3.39 : Self-Test Invalid
If the conditions are met, the following message is displayed.
Figure 3.40 : Confirm Self-Test
Leak Time: 3.45
Leak
Leak
Leak
4
±
8
000
Hide/Show Ps2
AC
92%
Start: Pt: 24.0000 inHg Ps2: 23.0000 inHg Ps1: 22.0000 inHg
( 1 min) 24.0022 0.0022 23.0021 0.0021 22.0024 0.0024
( 2 min) 24.0041 0.0041 23.0040 0.0040 22.0043 0.0043
( 3 min) 24.0059 0.0059 23.0057 0.0057 22.0060 0.0060
Press
CANC
to exit
Confirm Self-Test
Open all 3 ports to ambient, then press
GO
. Press
CANC
to exit
Self-Test Invalid
To perform Self Test
All three outputs must be in Measure mode and close to Ground
Press
CANC
to exit