6600-NG User’s Manual
125-9110A
Page 37
Figure 3.17 : Pt Rate Target value
Use the keypad to enter the desired value in the field shown. To clear an entry, press
CLR
. To accept the
entry, press
ENTER
. To exit this screen without changing the Rate Target value, press
CANC
. If the
number of digits entered (either before or after the decimal) exceeds the maximum allowed for the
current units, a long beep will be heard.
If the entered value exceeds the programmed limits, the following message will appear.
Pt Rate target exceeds limit. Press
CLR
Note: When Pt is in pressure units, the Pt Rate target is internally set to be equal to the Ps1 Rate target
while the Ps1 output is ramping. Once the Ps1 output stops ramping, the commanded Pt Rate target
becomes effective.
Note: In pressure units, Pt-Rate targets can be set between 0 and 30 inHg/min. (or equivalent)
If the Rate Target value needs to be changed in small increments, instead of entering a value, the arrow
keys may be used to Jog the Rate Target value up or down. If the current Pt Rate Target value is 150.0
knots, and the Lower screen appears as in Figure 3.17 above, pressing the
Up-Arrow
key will make the
screen appear as shown in Figure 3.18 below.
Figure 3.18 : Jog the Pt Rate Target value
3.9
20.0
Pt inHg
Pt
Ps2
Ps1
Measure
knots/min
0.0
151.0
inHg
inHg
29.9213
29.9213
knots
30.0000
30.0000
inHg/min
inHg/min
0.000
0.000
4.000
4.000
Measure
Measure
AC
92%
Hide/Show Ps2
000
Messages
Pt Rate Target
Enter Pt Rate target value: knots/min
Use keys to jog target value
Press
CANC
to exit