6600-NG User’s Manual
125-9110A
Page 43
Figure 3.25 : Pt Control mode
‘Control’ will be blinking in the Pt mode field. Also, ‘Control’ will be blinking in the Ps1 mode field. This
is because when either Pt mode, Ps2 mode or Ps1 mode is changed to Control, all three modes (Pt, Ps2,
Ps1) are automatically setup for Control. (
Caution:
Do NOT press GO at this moment
). Pressing
GO
will activate all 3 outputs (Pt, Ps2, Ps1) into Control mode. Pressing
CANC
will revert all 3 modes modes
back to Measure. For now, press
CANC
and return Pt, Ps2 and Ps1 to Measure mode.
Note: When Pt mode is changed to Control, Ps2 and Ps1 modes are automatically changed to Control.
Similarly, when Ps2 or Ps1 mode is changed to Control, Pt mode is automatically changed to Control.
Note: Leak and Measure modes can be selected independently for Pt, Ps2 and Ps1.
Ps1 Mode
The Ps1 mode can be changed when the key# symbol
(8)
is displayed to the right of the Ps1 mode field.
Press key ‘
8
’ on the keypad and the Lower section of the display will appear as shown in Figure 3.26
below.
Figure 3.26 : Select Ps1 Mode
Press the key with the # (shown in blue) associated with the mode to be selected. Press ‘
2
’ to select Leak
mode. The screen appears as shown in Figure 3.27 below.
Ps1 Mode
1)
Measure
2)
Leak
3)
Control
Press
CANC
to exit
Measure
Measure
Measure
000
Hide/Show Ps2
AC
92%
Messages
Control
Control
Control
4
±
8
000
Hide/Show Ps2
AC
92%