6600-NG User’s Manual
125-9110A
Page 41
Figure 3.22 : Jog the Ps2 Rate Target value
In this screen, the Upper section has a new symbol (with Up & Down arrows) to the left of the Ps2 Rate
Target value field. This indicates that the Ps2 Rate Target value can be jogged up or down repeatedly, in
steps of 0.01 inHg/min, using the Arrow keys, until
CANC
is pressed. In this case, since the
Up-Arrow
key was pressed, the Ps2 Rate Target value was incremented to 4.010 from 4.000 inHg/min. Press
Up-
Arrow
one more time and the value will change to 4.020. Press
Down-Arrow
two times and the value
decrements to 4.000.
Note: It is not possible to Jog a Rate Target value if it is blinking.
The incremental value for the Jog feature varies depending on Ps2 Rate units and is shown in the table
below.
Ps2-Rate Units
Jog increment
inHg/min
0.01
mbar /min
0.1
psi/min
0.01
mmHg/min
0.1
kpa/min
0.01
Feet/min
10
Meters/sec
0.1
km/hr/sec
0.1
29.9213
30.0000
Pt inHg
Pt
Ps2
Ps1
Measure
inHg/min
0.000
4.000
Dfin
inHg
0.0001
29.9213
Pt inHg
0.1010
30.0000
inHg/min
inHg/min
0.000
0.000
4.010
4.000
Measure
Measure
AC
92%