Test Procedures
APPLICATION:
Swimming pool water.
RANGE:
0–150 Cyanuric Acid
MDL:
5 ppm
METHOD:
A buff ered solution of melamine forms a precipitate with
cyanuric acid in proportion to the amount of cyanuric acid
present. The amount of particles in suspension is measured
turbidimetrically.
SAMPLE HANDLING &
PRESERVATION:
Cyanuric acid samples should be analyzed as soon as
possible after collection. Deterioration of the sample can be
minimized by keeping samples in the dark or refrigerated
until analysis can be performed.
INTERFERENCES:
No known interference from compounds normally found in
pool water. Temperature of the sample should be maintained
between 70°F and 80°F for best results. Check for stray light
interference (see page 16).
CYANURIC ACID
, UDV
SMART Spectro 2 Test Procedures 05.02.16