84
SMART Spectro 2 Operator’s Manual 04.16
WARNING SIGNAL TOO STRONG
This can be an indication that the meter was accidently shut-off during wavelength
calibration. Shut meter off and restart.
WARNING RAM BATTERY LOW
Clock battery needs to be replaced soon. If it fails all user tests and Data log Data will be
lost. Unit can be run without this battery. Just use line power and leave unit on.
HELPFUL HINTS
POWER
The power supply has an internal switch that handles both 110V and 220V input.
STRAY LIGHT
The SMART Spectro 2 should have no problems with stray light. Make sure that the sample
compartment lid is always fully closed.
ACCIDENTAL LOSS OF POWER
If for any reason the meter experiences a loss of power during wavelength calibration,
the next time the meter is powered up a wavelength calibration will automatically be
performed.
DEFAULT SETTINGS
To reset the SMART Spectro 2 to the default settings, see RESET TO DEFAULT (page 43).