Test Procedures
CHLORIDE, TESTAB
ARGENTOMETRIC TESTAB METHOD • CODE 3693-SC
QUANTITY
CONTENTS
CODE
50
*Chloride Spectrophotometric Grade Tablets
*3885A-H
1
Tablet Crusher
0175
*WARNING:
Reagents marked with an * are considered to be potential health hazards. To
view or print a Safety Data Sheet (SDS) for these reagents go to www.lamotte.com.
Search for the four digit reagent code number listed on the reagent label, in the contents
list or in the test procedures. Omit any letter that follows or precedes the four digit code
number. For example, if the code is 4450WT-H, search 4450. To obtain a printed copy,
contact LaMotte by email, phone or fax.
Emergency information for all LaMotte reagents is available from Chem-Tel:
(US, 1-800-255-3924) (International, call collect, 813-248-0585).
Chloride is one of the major anions found in water and sewage. The presence of chlorides in
large amounts may be due to the natural process of water passing through salt formations
in the earth, or it may be evidence of the intrusion of seawater or pollution from industrial
processes or domestic wastes. The salt content of water aff ects the distribution of plant
and animal life in an aquatic system, based on the amount of salt they can tolerate.
APPLICATION:
Drinking, surface, and saline waters; domestic and industrial
wastewaters.
RANGE:
0.0–30.0 ppm Chloride
MDL:
0.5 ppm
METHOD:
Silver nitrate reacts with chloride to form turbid silver chloride
in proportion to the amount of chloride in the sample.
SAMPLE HANDLING &
PRESERVATION:
Collect samples in clean, chemically-resistant glass or plastic
containers. No preservative is needed if sample is to be stored.
INTERFERENCES:
Substances in amounts normally found in drinking water will
not interfere. Bromide, iodide, cyanide, sulfi de, thiosulfate,