Test Procedures
COLLECTION & TREATMENT OF THE WATER SAMPLE
Steps 1 through 4 below describe proper sampling technique in shallow water. For sample
collection at depths beyond arm’s reach, special water sampling apparatus is required
(e.g. the LaMotte Water Sampling Chamber, Code 1060; Model JT-1 Water Samplers, Code
1077; Water Sampling Outfi t, Code 3103; or Water Sampling Bottle, Code 3-0026).
1. To avoid contamination, thoroughly rinse the screw cap Sample Tube (29180) with
sample water.
2. Tightly cap Sample Tube and submerge to the desired depth. Remove cap and allow
the Sample Tube to fi ll.
3. Tap the sides of the submerged tube to dislodge any air bubbles clinging to the inside.
Replace the cap while the Sample Tube is still submerged.
4. Retrieve Sample Tube and examine it carefully to make sure that no air bubbles are
trapped inside. Once a satisfactory sample has been collected, proceed immediately
with Steps 5 and 6 to “fi x” the sample.
NOTE: Be careful not to introduce air into the sample while adding the reagents
in steps 5 and 6. Simply drop the reagents into the sample. Cap carefully, and mix
gently.
5. Add 2 drops of *Manganese Sulfate Solution (4167) and 2 drops of *Alkaline
Potassium Iodide Azide (7166). Cap and mix by inverting several times. A precipitate
will form. Allow the precipitate to settle below the shoulder of the tube before
proceeding.
6. Add 8 drops of *Sulfuric Acid, 1:1 (6141WT). Cap and gently mix until the reagent
and the precipitate have dissolved. A clear-yellow to brown-orange color will develop,
depending on the oxygen content of the sample.
NOTE: It is very important that all “brown fl akes” are dissolved completely. If the water has a
high DO level this could take several minutes. If fl akes are not completely dissolved after 5
minutes, add 2 drops of *Sulfuric Acid 1:1 (6141WT) and continue mixing.
Following the completion of step 6, contact between the water sample and the
atmosphere will not aff ect the test result. Once the sample has been “fi xed” in this manner,
it is not necessary to perform the actual test procedure immediately. Thus, several
samples can be collected and “fi xed” in the fi eld, and then carried back to a testing station
or laboratory where the test procedure is to be performed.