Test Procedures
APPLICATION:
Drinking and surface water and swimming pool water
RANGE:
0–200 ppm as CaCO
3
MDL:
15 ppm
METHOD:
The sample is added to a buff ered indicator reagent. The color
that develops, ranging from yellow to blue, will indicate the
amount of alkalinity in the sample.
SAMPLE HANDLING &
PRESERVATION:
Samples should be analyzed as soon as possible after
collection. Sample may be refrigerated for 24 hours.
INTERFERENCES:
Quats and poly quats at high concentrations will interfere.
ALKALINITY
, UDV
SMART Spectro 2 Test Procedures 05.02.16