SMART Spectro 2 Operator’s Manual 04.16
CALIBRATE WL
The Calibrate Wavelength (CALIBRATE WL) mode is used to establish or re-establish
the accuracy of the wavelength selection process. Normally, the Calibrate Wavelength
procedure should be run after the SMART Spectro 2 is turned ON and allowed to warm
up for 15 minutes or if operating conditions (temperature, humidity, etc.) change
signifi cantly.
For fi eld use, when operating with the battery, calibrate wavelength prior to going into
the fi eld using AC power. This will increase battery life in the fi eld. Alternatively calibrate
wavelength in the fi eld immediately before testing. Turn Spectro 2 on immediately before
scanning blank. Calibrate wavelength just before scanning blank.
1. Press
ON
. The LaMotte logo screen
will appear for about 2 seconds. After
initialization the MAIN MENU will appear.
MAIN MENU 12:00:00
PROGRAMMED TESTS
%T/ABS
PC LINK
SYSTEM SETUP
2. Press or
to scroll to
SYSTEM
SETUP
.
MAIN MENU 12:00:00
PROGRAMMED TESTS
%T/ABS
PC LINK
SYSTEM SETUP