Test Procedures
PROCEDURE
Use universal sample holder.
1. Press
ON
button to turn spectrophotometer on.
2. Scroll to and select
PROGRAMMED TEST
from menu.
3. Scroll to and select
ALL TESTS
(or another sequence containing
29 Copper BCA-LR
) from
TESTING MENU
.
4. Scroll to and select
29 Copper BCA-LR
from menu.
5. Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with sample.
6. Insert tube into chamber, close lid and select
SCAN BLANK
.
7.
Remove tube from Spectro and add one *Copper Tablet (3808A). Cap and shake
vigorously until tablet dissolves. Solution will turn purple if copper is present. Wait 2
minutes.
8. At end of 2 minute waiting period, mix, insert tube into chamber, close lid and select
SCAN SAMPLE
. Record result.
9. Press and hold
OFF
button to turn spectrophotometer off or press
EXIT
button to exit
to a previous menu or make another menu selection.
COPPER
, LOW RANGE
SMART Spectro 2 Test Procedures 05.02.16