SMART Spectro 2 Operator’s Manual 04.16
67
10. Press
*ENTER
to move to the second
standard. Use the same procedure to
scan the blank and then scan the second
standard to create a second data point.
Repeat the procedure until the data from
all of the standards has been entered into
the calibration.
NEW CALIBRATION 465NM
ENTER STD. 1
ENTER STD. 2
ENTER STD. 3
ENTER STD. 4
CALCULATE
NOTE: To escape from
NEW CALIBRATION
before completing a standard selection, select
blank and press
*ENTER
. After blanking, scroll to
!
and press
*ENTER
. Then press
EXIT
.
11. Press
*ENTER
to select
CALCULATE
.
SELECT DEGREES 465NM
1. DEGREE THRU.0
1. DEGREE
2. DEGREE
3. DEGREE
12. SCROLL
or
to select the desired
curve fi t.
SELECT DEGREES 465NM
1. DEGREE THRU 0
1. DEGREE
2. DEGREE
3. DEGREE
Note:
1. DEGREE THRU.0
calculates the best straight line fi t through the data points and
intersects with the origin at 0 ppm, 0 absorbance. This is a classical Beers Law calibration.
If a one point calibration has been performed, the
1. DEGREE THRU.0
curve fi t must be