SMART Spectro 2 Operator’s Manual 04.16
65
4. Press
*ENTER
to save the selection.
NEW CALIBRATION 465NM
ENTER STD. 1
ENTER STD. 2
ENTER STD. 3
ENTER STD. 4
CALCULATE
5. Press
*ENTER
to select the fi rst data
point.
ENTER STD. 1 465NM
SCAN BLANK
STD. 1 =
! 0 1 2 3 4 5 6 7 8 9 .
6. Place a blank in the sample chamber.
Close lid. Press
*ENTER
to
SCAN BLANK
.
The display will indicate that the blank is
being scanned. The display will indicate
BLANKED. A highlighted box will appear on
the character selection line.
ENTER STD. 1 465NM
BLANKED
STD. 1 =
! 0 1 2 3 4 5 6 7 8 9 .