SMART Spectro 2 Operator’s Manual 04.16
7.
Place the reacted standard of the lowest
concentration in the chamber. Press
or
to highlight each character in the
concentration value. Press
*ENTER
after
each selection to confi rm the selection.
ENTER STD. 1 465NM
BLANKED
STD. 1 =
! 0 1 2 3 4 5 6 7 8 9 .
8. After the concentration has been entered,
scroll to
!
. Press
*ENTER
to save the value
and scan the standard. The absorbance
will be displayed.
ENTER STD. 1 465NM
BLANKED
STD. 1 = 1.0
A = 0.4116
9. Press
EXIT
to return to the New Calibration
menu.
NEW CALIBRATION 465NM
ENTER STD. 1
ENTER STD. 2
ENTER STD. 3
ENTER STD. 4
CALCULATE