- 25 -
GB
PROBLEM
POSSIBLE CAUSE
REMEDY
blinking
The built-in temperature overload sensor
is defective.
Contact Kulzer or local authorised
Customer Service agent.
Note:
If the temperature overload sensor
fault occurs during an exposure cycle,
the software allows completing the
exposure. No further exposure cycle is
allowed by pressing the “
STAND-BY
”
button and “
4 beeps
” are emitted.
12 Technical
specifi cations
Classifi cation according to Directive 93/42/EEC:
Class I (one)
Safety Standards:
The product has been tested and found to comply to IEC 60601-1
(second edition) and to IEC 60601-1-1.
The product has been tested and found to comply to IEC 60601-1
(third edition) +AMD1:2012 and to IEC 60601-1-2:2014 (fourth
edition – for electro-magnetic compatibility – EMC).
The hand piece is intended for use within the patient environment
whereas the charger unit and the power supply adapter are not
intended for use within the patient environment.
Note:
The patient environment is defi ned as an area 1.5 m from the
patient as per IEC 60601-1 (third edition) +AMD1:2012.
Classifi cation according to IEC 60601-1:
Class II, Type B,
Applied Part:
Light guide with protective cone.
IP 20 (Charging unit).
IP 20 (Translux 2Wave)
Charging unit:
Translux 2Wave Charging unit
Power supply for the charging unit:
Switching Power Supply (Direct Plug-In) manufactured by:
Phihong Technology CO LTD –
Model:
PSACO5R-050L6.
Input:
100 – 240 V
– 300 mA – 50 / 60 Hz –.
Output:
5 V
– 1 A Max –
.
Class II equipment
Power supply for the hand piece:
Rechargeable Lithium-ion battery.
Manufacturer:
Panasonic.
Model:
NCR-18500.
Nominal Voltage:
3.6 V.
Nominal capacity
(Typical): 2000 mAh
Hand piece:
Model:
Translux 2Wave.
Operation: Intermittent service:
120“ ON – 40“ OFF –
max 2 times running
Light source:
High-luminosity
LED.
LED Product Risk Group 2 (IEC 62471).
Wavelength range:
385 – 510 nm.
Dominant wavelength:
Min. 385 nm – Max. 470 nm
Light guide:
Lead-free multi core rod, 8 mm diameter.
Sterilisable in pre-vacuum steam autoclave:
4 minutes (minimum) at minimum temperature of 132°C + 3°C
(270°F + 5°F). Max. 500 cycles
Settable exposure cycles:
– “Soft-Start” exposure
20 seconds of exposure time. (The “
Soft-Start
” means increase of
light intensity from 50% up to 100% within 2 seconds.)
Acoustical signals:
1 beep
at the start of exposure.
1 beep
after 10 seconds of exposure.
1 beep
at the end of exposure
– Normal exposure
20, 10 or 5 seconds of exposure time.
Acoustical signals:
1 beep
at the start of exposure.
1 beep
after 10 seconds of exposure (only 20 s Mode).
1 beep
at the end of exposure
1707_11667_GBA_Translux_2Wave_EU_NENA.indb GB25
1707_11667_GBA_Translux_2Wave_EU_NENA.indb GB25
07.09.17 16:22
07.09.17 16:22