Chapter 3 Troubleshooting
3-34
3.3.2 Power ON/OFF Related
Phenomenon that might occur when the power is turned ON/OFF is described here.
There are REGIUS console and JM with same
IP address and host name.
•
When you set the IP address and host name of
the REGIUS console to the one used in the JM,
REGIUS console will not startup.
•
When you set the IP address and host name of
the JM to the one used in the REGIUS console, it
will read normally.
There are REGIUS MODEL 110 and REGIUS
console with same host name.
•
Even if you set the host name of the REGIUS
console as same as the REGIUS MODEL 110, it
will read as normal.
•
Even if you set the host name of the REGIUS
console as same as the REGIUS MODEL 110, it
will read as normal.
There are REGIUS MODEL 110 and JM with
same host name.
Either way, it will read normally.
Phenomenon
Response
Nothing operates when "operation"
switch is pressed.
("operation" lamp does not turn on)
Analysis Procedure of Power Supply Problems
(Cannot Turn ON the Power Supply) (Page 3-35)"
When the "operation" switch is pressed
for 3 seconds to shutdown during the
"READY" state, message display
window turns off but the blinking of the
LED does not stop.
Turn OFF the power supply circuit breaker of the equipment.
"operation" switch was pressed before
the JM started up.
Error (24000) will occur, but it will connect automatically as
soon as the JM starts up, and read will be possible.
"operation" switch was pressed when
there was an error, but it did not start the
power OFF action.
Turn OFF the power supply circuit breaker of the equipment.
Pressed the "operation" switch during
the read.
•
Image getting read will be sent to REGIUS console.
•
Equipment will shutdown automatically after the transfer
of the image. Startup the equipment when the
"operation" switch was pressed by mistake.
Sleep action does not function properly.
Confirm the Relation setting in the JM.
Equipment does not startup after starting
up the REGIUS console.
Confirm if the "operation" lamp is on (sleep mode).
•
If the "operation" lamp is off, power supply is OFF
(circuit breaker OFF) or stopped, so turn ON the power
supply circuit breaker and startup by pressing the
"operation" switch.
•
If the "operation" lamp is on, shutdown once by pressing
the "operation" switch and restart the equipment.
Setup condition
Phenomenon
Содержание 0902
Страница 2: ......
Страница 6: ......
Страница 12: ......
Страница 30: ...2 10...
Страница 54: ......
Страница 84: ...Chapter 3 Troubleshooting 3 30 3 2 19 Confirm the H sync Signal To be stated...
Страница 85: ...Chapter 3 Troubleshooting 3 31 3 2 20 Confirm the LMC Operation To be stated...
Страница 120: ......
Страница 296: ...8 10...
Страница 300: ......
Страница 301: ......