12002 - EXTRACTION BAR REFERENCE SENSOR S9 failed to activate
12003 - PLATE PRESENT SENSOR S5 did not detect PLATE
Cause 1: The EXTRACTION BAR SENSOR S9 operates correctly, but another malfunction
occurred.
Action:
1.
Remove any obstructions from the path of the EXTRACTION BAR.
2.
Do the checkout for the
SLOW SCAN CONTROLLER BOARD A6 and COIL
Cause 2: The EXTRACTION BAR SENSOR S9 malfunctioned.
Action:
Do the checkout for the
Cause 1: The PLATE PRESENT SENSOR S5 operates correctly, but another malfunction
occurred.
Action:
1.
Check:
•
PLATE was removed from the CASSETTE
•
SLOW SCAN MOTOR operates correctly
2.
Do the checkout for the
SLOW SCAN CONTROLLER BOARD A6 and COIL
3.
Remove any obstructions from the path of the EXTRACTION BAR.
Cause 2: The PLATE PRESENT SENSOR S5 malfunctioned.
Action: